For other PSI sensors, select a sensor from the menu below:
The CO2-Pro Atmosphere™ is an
optional features for the CO2-Pro™ that provides measurements of
pCO2 in water and in the Atmosphere using the same detector, which
means that any calibration offset is eliminated in the calculation
of ΔpCO2. The CO2-Pro™
has a patented, revolutionary new interface that provides an
equilibrated gas sample to the detector. Note that a source of water
flow is required at 3 to 10 litres per minute (optional Sea-Bird 5-T
pump available from Pro-Oceanus Systems Inc. or from Sea-Bird
Electronics, Inc.).
The CO2-Pro™
is factory calibrated from 0-600 ppm CO2 with calibration burned
into EPROM (other calibration ranges are available by special
order). To maintain accuracy, the detector module is equipped to
perform an automatic zero point calibration (AZPC). When the AZPC is
initiated, the gas stream is routed through a CO2 absorbent to
provide a zero ppm CO2 measurement. The CO2 zero measurement
compensates for changes in optical cell performance and is used in
determinations of ppm CO2 until a new AZPC is performed.
PSI CO2-ProTM Features:
High precision and accuracy
RS-232 data interface
Low power requirements
Light weight and compact size
Unique antifouling protection
Raw data and ppm CO2 included in data stream
Low cost
Comes equipped with the PSI patented tubular pumped fast interface.
Download the CO2-Pro™ Atmosphere™ Product Brochure
HERE
Physical Specifications
Length: 33 cm (43 cm with connectors)
Diameter: 17.3 cm
Weight (air): 19.8 lbs. (9 kg)
Weight (water): 0.88 lbs. (0.4 kg)
Housing:
Hard anodized aluminum
Proof to 1000 meters (Contact PSI for details)
Electrical Characteristics
Input voltage:
12 ± 0.5 Volts DC through 4-pin impulse male connector
With power-saving option: 9.5-18 Volt DC
Power consumption (not including pump power):
During warm-up: 0.8 A at 12 volts
During operation: 0.4 A
Lower power consumption options available
Data storage:
With Data Logger: 2 mega-bytes EPROM
Data output:
RS-232 serial output
Sample Rate:
0.6 hertz without controller; 0.3 hertz in continuous mode with
controller
Equilibration time constant:
Approximately 3 minutes (depending on pump flow rate)
Performance
Accuracy
CO2 concentration ±~ 1 ppm
Gas stream humidity ± 1 mb
Gas stream pressure ± 2 mb
Precision
CO2 concentration 0.01 ppm
Gas stream humidity 1 mb
Gas stream pressure 1 mb
Calibration range 0-600 ppm (other ranges available by special
order)
Temperature range 0-35°C